| Natural¹ | Induced (Storage and Transit) | |||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| High Humidity | Hot Humid (Cycle B3) | Constant Temp. (Cycle B1) | Cyclic High RH (Cycle B2) | Hot Humid (Cycle B3) | ||||||||||||||
| Constant Temp. (Cycle B1) | Cyclic High RH (Cycle B2) | |||||||||||||||||
| Temp | RH | Temp | RH | Temp | RH | Temp | RH | Temp | RH | Temp | RH | |||||||
| Time | °C | °F | % | °C | °F | % | °C | °F | % | °C | °F | % | °C | °F | % | °C | °F | % |
| 0100 | Nearly constant at 24°C (75°F) throughout the 24 hours | 27 | 80 | 1002 | 31 | 88 | 88 | Nearly constant at 27°C (80°F) throughout the 24 hours | 33 | 91 | 69 | 35 | 95 | 67 | ||||
| 0200 | 26 | 79 | 100 | 31 | 88 | 88 | 32 | 90 | 70 | 34 | 94 | 72 | ||||||
| 0300 | 26 | 79 | 100 | 31 | 88 | 88 | 32 | 90 | 71 | 34 | 94 | 75 | ||||||
| 0400 | 26 | 79 | 100 | 31 | 88 | 88 | 31 | 88 | 72 | 34 | 93 | 77 | ||||||
| 0500 | 26 | 78 | 100 | 31 | 88 | 88 | 30 | 86 | 74 | 33 | 92 | 79 | ||||||
| 0600 | 29 | 78 | 100 | 32 | 90 | 85 | 31 | 88 | 75 | 33 | 91 | 80 | ||||||
| 0700 | 27 | 81 | 94 | 34 | 93 | 80 | 34 | 93 | 64 | 36 | 97 | 70 | ||||||
| 0800 | 29 | 84 | 88 | 36 | 96 | 76 | 38 | 101 | 54 | 40 | 104 | 54 | ||||||
| 0900 | 31 | 87 | 82 | 37 | 98 | 73 | 42 | 107 | 43 | 44 | 111 | 42 | ||||||
| 1000 | 32 | 89 | 79 | 38 | 100 | 69 | 45 | 113 | 36 | 51 | 124 | 31 | ||||||
| 1100 | 33 | 92 | 77 | 39 | 102 | 65 | 51 | 124 | 29 | 57 | 135 | 24 | ||||||
| 1200 | 34 | 94 | 75 | 40 | 104 | 62 | 57 | 134 | 22 | 62 | 144 | 17 | ||||||
| 1300 | 34 | 94 | 74 | 41 | 105 | 59 | 61 | 142 | 21 | 66 | 151 | 16 | ||||||
| 1400 | 35 | 95 | 74 | 41 | 105 | 59 | 63 | 145 | 20 | 69 | 156 | 15 | ||||||
| 1500 | 35 | 95 | 74 | 41 | 105 | 59 | 63 | 145 | 19 | 71 | 160 | 14 | ||||||
| 1600 | 34 | 93 | 76 | 41 | 105 | 59 | 62 | 144 | 20 | 69 | 156 | 16 | ||||||
| 1700 | 33 | 92 | 79 | 39 | 102 | 65 | 60 | 140 | 21 | 66 | 151 | 18 | ||||||
| 1800 | 32 | 90 | 82 | 37 | 99 | 69 | 57 | 134 | 22 | 63 | 145 | 21 | ||||||
| 1900 | 31 | 88 | 86 | 36 | 97 | 73 | 50 | 122 | 32 | 58 | 136 | 29 | ||||||
| 2000 | 29 | 85 | 91 | 34 | 94 | 79 | 44 | 111 | 43 | 50 | 122 | 41 | ||||||
| 2100 | 28 | 83 | 95 | 33 | 91 | 85 | 38 | 101 | 54 | 41 | 105 | 53 | ||||||
| 2200 | 28 | 82 | 96 | 32 | 90 | 85 | 35 | 95 | 59 | 39 | 103 | 58 | ||||||
| 2300 | 27 | 81 | 100 | 32 | 89 | 88 | 34 | 93 | 63 | 37 | 99 | 62 | ||||||
| 2400 | 27 | 80 | 100 | 31 | 88 | 88 | 33 | 91 | 68 | 35 | 95 | 63 | ||||||
507.6 -II. Test cycles (days)
| MATERIEL CATEGORY | NATURAL¹ | INDUCED (STORAGE and TRANSIT) | ||||
|---|---|---|---|---|---|---|
| Cycle B1 | Cycle B2 | Cycle B3 | Cycle B1 | Cycle B2 | Cycle B3 | |
| Hazardous Items Normal Test Duration | 90 | 90 | 30 | 180 | 180 | 30 |
| Non-Hazardous Items Normal Test Duration | 45 | 45 | 15 | 90 | 90 | 15 |
¹ Perform operational checks at least once every five days; more frequent checks may provide early detection of potential problems.
- c. Procedure II – Aggravated Cycle. For Procedure II, in addition to a 24-hour conditioning cycle (paragraph 4.4.2.2), the minimum number of 24-hour cycles for the test is ten. Although the combined 60 °C (140 °F) and 95 percent RH does not occur in nature, this combination of temperature and relative humidity, has historically proven adequate to reveal potential effects in most materiel. Extend the test as specified in the test plan to provide a higher degree of confidence in the materiel to withstand warm, humid conditions. For the test items incorporating seals to protect moisture-sensitive materials, e.g., pyrotechnics, longer test durations may be required.